| Model Number |
3P36-35 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer stated that a falsely elevated architect afp result of 29.83 ng/ml was generated for a patient sample on (b)(6) 2022.The same sample tube was retested on (b)(6) 2022 and the result was 2.53 ng/ml.The sample was tested on a different architect which confirmed the lower result (2.6 and 2.67 ng/ml).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.Worldwide data from abbottlink was reviewed and determined that patient median result for the lot is comparable with other lots in the field and within established baselines confirming no systemic issue for the lot.Based on the investigation, no deficiency for lot number 33052fn00 was identified.
|
| |
|