• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52917
Medical Device Problem Codes Obstruction of Flow (2423); Structural Problem (2506); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Internal Organ Perforation (1987); Perforation of Vessels (2135); Thrombosis/Thrombus (4440); Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter occupation: non-healthcare professional.(510k): k211874.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.20 devices in lot.No relevant notes on wo for neither device (igtcfs-65-1-fem-tulip) lot: e2995015, nor filter (igtf-30) lot: e2991536 and e2991506.Three other complaints ((b)(4), different issue, occasional cramps on right side, pr361557, different issue, ivc filter perforation ivc unfiled claims and (b)(4), different issue, ivc filter perforation ivc unfiled claims) on lots.Product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It is alleged that the patient received a gunther tulip inferior vena cava (ivc) filter on (b)(6) 2013, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: a2, a4, b1, b5, b6, b7, h6.Investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: organ/ vena cava (vc)/ vertebra perforation, deep vein thrombosis (dvt), fear.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Unknown if the reported fear is directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.To date, three (3) other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Patient allegedly received an implant on (b)(6) 2013 via the right common femoral vein due to pulmonary emboli (pe).Patient is alleging organ and vena cava perforation, vertebral perforation.Patient further alleges "fear that the filter might cause further damage as it has not yet been removed" and deep vein thrombosis of left leg after filter placement.Per ct report, "3 mm aortic, 5 mm vertebral, and 4 mm mesenteric perforation.No fracture, tilting, migration, or stenosis".Per ct report, "an ivc filter was identified.The tip of the filter is not in contact with the wall of the ivc.The metallic struts appeared to be intact.Only 1 of the struts perforates the wall of the ivc by strict definition.It is the strut in the 4 o'clock position and it perforates the wall by approximately 1.5 mm.The struts do not extend into any vital structures.The superior tip of the ivc filter is 2.4 cm below the right and left renal vein orifices.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key14846983
Report Number3002808486-2022-00864
Device Sequence Number4640039
Product Code DTK
UDI-Device Identifier10827002529172
UDI-Public(01)10827002529172(17)151018(10)E2995015
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2015
Device Model NumberG52917
Device Catalogue NumberIGTCFS-65-1-FEM-TULIP
Device Lot NumberE2995015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexFemale
Patient Weight97 KG
-
-