| Model Number |
7132 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.Vascular access was obtained via the left common femoral artery.The 70% stenosed target lesion was located in the mildly tortuous and moderately calcified vessel.The 2.50mm x 12mm emerge balloon was introduced but became lodged in the lesion.The device could not be advanced and when removal was attempted, the shaft broke.The fractured balloon shaft was snared and removed.The procedure was completed with another balloon.There were no patient complications.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.Vascular access was obtained via the left common femoral artery.The 70% stenosed target lesion was located in the mildly tortuous and moderately calcified vessel.The 2.50mm x 12mm emerge balloon was introduced but became lodged in the lesion.The device could not be advanced and when removal was attempted, the shaft broke.The fractured balloon shaft was snared and removed.The procedure was completed with another balloon.There were no patient complications.It was further reported via facility medwatch report mw5110587 that while the balloon was in the distal aorta, the shaft broke.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event: corrected.
|
| |
|