Model Number US339
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Superficial (First Degree) Burn (2685)
Date of Event 05/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation on-going.Should relevant additional information/investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap inc.That an adson bipolar forceps 1.0mm127mm (part # us339) was used during a procedure performed on (b)(6) 2022.According to the complainant, during the procedure, the patient was burned on the cheek while using bipolar.The complaint device was returned to the manufacturer for evaluation.The burn was treated with ointment.Although requested, additional information has not been made available.The malfunction is filed under aic reference (b)(4).
 
Additional Manufacturer Narrative
Manufacturing site evaluation: the complaint device was returned to the manufacturer for evaluation.A visual examination revealed that the tips of the device were charred and pitted, while the rest of the device appeared to be in good condition.No other damage was visible.The device history records (dhr) were reviewed for the provided serial number; the device was found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.There was no evidence to suggest any defect in the design, materials, or workmanship of the device.Based on the investigation results, the most probable root cause was determined to be operational error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADSON BIPOLAR FORCEPS 1.0MM127MM
Common Device Name
BIPOLAR COAGULATION FORCEPS
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
jonathan severino
3773 corporate parkway
center valley, PA 18034
4847197287
MDR Report Key14847227
Report Number2916714-2022-00068
Device Sequence Number2382053
Product Code GEI
UDI-Device Identifier04046964582417
UDI-Public4046964582417
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K811820
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberUS339
Device Catalogue NumberUS339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-