MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202901
Device Problem Break (1069)
Patient Problems Perforation (2001); Foreign Body In Patient (2687); Unspecified Tissue Injury (4559)
Event Date 06/08/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that during a shoulder cuff repair, the footprint anchor broke.The procedure was completed with a non-significant delay using a back-up device in an additional bone hole.Pieces were not removed from the patient.No further complications were reported.Patient status is good.
 
Manufacturer Narrative
Internal complaint reference (b)(4).H10 h3, h6: the reported device was received for evaluation.There was a relationship found between the returned device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A visual inspection found the device outside the original packaging.The insertion device has debris on it, the distal end is bent.The suture threads are still through the suture window in the anchor and the anchor is fractured at the proximal end.There is debris on the anchor and suture material.Based on the condition of the product material found during visual inspection, additional material testing is not required.This case reports the breakage of the footprint anchor and per case details the broken pieces were not retrieved from the patient.No clinically relevant supporting documentation was provided for inclusion in a medical investigation.The footprint anchor is implantable, biocompatibility is not an issue.Micro-motion or movement cannot be predicted.There is potential risk for tissue inflammation and/or pain.It was also reported that an additional bone hole was required to complete the procedure and the impact to the patient is expected to be minimal.No further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14847304
MDR Text Key294949127
Report Number1219602-2022-00925
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010650528
UDI-Public03596010650528
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K093897
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202901
Device Catalogue Number72202901
Device Lot Number2069046
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received08/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-