• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52917
Medical Device Problem Codes Structural Problem (2506); Device Tipped Over (2589)
Health Effect - Clinical Code Perforation of Vessels (2135)
Type of Reportable Event Serious Injury
Event or Problem Description
The following information is alleged: the patient received a gunther tulip inferior vena cava (ivc) filter infrarenally on (b)(6) 2014 to avoid the risk of pulmonary embolism.Approximately 4 years and 11 months later, the patient underwent a computerized tomography (ct) scan of the patient's abdomen.The ct scan revealed that the struts of the cook tulip filter had perforated outside of the wall of the ivc.Hospital and medical records have been requested, but not yet provided.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Non-healthcare professional.(510k): (b)(4).Investigation it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.The following allegations have been investigated: vena cava perforation.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: vascular trauma, vena cava perforation, vena cava penetration.20 devices in lot.No relevant notes on wo for neither device (igtcfs-65-1-fem-tulip) lot: e3206401, nor filter (igtf-30) lot: e3202180 and e3199607.No other complaints on lots.Product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Patient allegedly received the gunther tulip filter via the right common femoral vein.Per a computed tomography (ct) abdomen: "findings: the apex of the ivc filter is at the level of the renal veins.The struts are well below the level of the renal veins and renal arteries.There is slight leftward tilt of the apex of the filter, the strut in the 2 o'clock position is perforating the wall of the ivc by approximately 2 mm.The strut at the 4 o'clock position similarly perforates the ivc by approximately 2 mm.Smaller struts which are more proximal in location do not perforate the vessel.There is no extension of struts into any of the adjacent organs".
 
Additional Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: tilt.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.20 devices in lot.No other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key14847437
Report Number3002808486-2022-00866
Device Sequence Number4640048
Product Code DTK
UDI-Device Identifier10827002529172
UDI-Public(01)10827002529172(17)170416(10)E3206401
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2017
Device Model NumberG52917
Device Catalogue NumberIGTCFS-65-1-FEM-TULIP
Device Lot NumberE3206401
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer03/01/2023
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date03/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/01/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexMale
-
-