Model Number 3500B
Device Problem Mechanical Problem (1384)
Patient Problem Low Oxygen Saturation (2477)
Event Date 03/08/2019
Event Type  malfunction  
Event Description
This report is being submitted in response to voluntary mdr report no.Mw5084865 found in the maude database.Sechrist had no prior knowledge of this incident."it was reported that pt on.5ml of sweep prior to intubation increased flow and sweep for intubation.After intubation pt's o2 sats were stable; orders from physician to put sweep at.5 and wean to cap as tolerated.Pt capped and abg pco2 elevated.Rechecked abg to verify results.When i went to increase sweep gas blender knob was difficult to turn.Sweep placed at 6l.".
 
Manufacturer Narrative
This report is being submitted in response to voluntary mdr report found in the maude database.Sechrist had no prior knowledge of this incident.This report is based solely on the customer's reported issue.Based on the age of the device, it is unlikely that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file no.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECHRIST AIR/OXYGEN GAS BLENDER MODEL 3500CP-G
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer (Section G)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer Contact
majid mashayekh
4225 e. la palma ave.
anaheim, CA 92807
7145798309
MDR Report Key14847537
MDR Text Key295136502
Report Number2020676-2022-00015
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3500B
Device Catalogue Number3500B
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/09/2022
Initial Date FDA Received06/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/1988
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-