• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC SUREFORM; STAPLER 60 RELOAD BLACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC SUREFORM; STAPLER 60 RELOAD BLACK Back to Search Results
Model Number 48360T-08
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 01/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the sureform 60 black reload accessory involved with this complaint and completed device evaluation.Failure analysis (fa) could not replicate/confirm the customer reported complaint.A visual inspection noted no damage was found.The reload was not fired.The sureform 60 stapler instrument involved with this complaint was also returned and completed device evaluation.Fa was able to confirm but not replicate the reported complaint.The instrument was found to have repeated clamping failures based on review, but was not able to be replicated in-house.The stapler instrument was installed a total of 8 times, and clamping failures occurred on installs 1, 2, and 7.It was noted the clamp history shows 10 incomplete clamps, 7 aborted, and 18 attempts.No image or procedure video was provided for review.A review of the system logs for the event date of (b)(6) 2020 has been performed by an isi post market surveillance specialist and the following was observed: there were no observed events in the system logs that would suggest a product issue and the logged events are in line with normal system functionality.This complaint is being reported due to the following conclusion: it was alleged that during a da vinci-assisted pulmonary segmentectomy surgical procedure, tissue was stuck to the stapler/reload after firing due to an unknown cause.As a result, the procedure was converted to traditional laparoscopic surgery.The patient experienced bleeding, which the surgeon resolved with electrocautery.Follow-up was attempted, but the missing patient information was either unknown, unavailable, not provided, or not applicable.The expiration date is not applicable.The product is not implantable.It is unknown if the initial reporter submitted a report to the fda.
 
Event or Problem Description
It was reported that during a da vinci-assisted pulmonary segmentectomy surgical procedure, the sureform 60 stapler indicated that the tissue was too thick, even after using the black stapler reloads.The target tissue was reported as lung and there was no tissue tension or tissue bunching during the stapler clamping sequence.The procedure was converted to traditional laparoscopic surgery.An error message: "tissue too large" was generated when the stapling issue occurred.Intuitive surgical, inc.(isi) obtained the following additional information from the site¿s robotics coordinator: the procedure was converted to traditional laparoscopic surgery because the issue occurred towards the end of the case and the surgeon didn't want to open up a new stapler instrument.The stapler successfully fired a few times, then would either fire successfully during the case or not fire at all.The stapler did not partially fire and there were no malformed staples.After firing, the patient's tissue would stick to the stapler.The surgeon used the stapler's manual release knob (mrk) to separate the stapler from the patient's tissue.Bleeding was observed during the procedure and blood loss volume was noted to be less than 750 ml and the surgeon used electrocautery to stop the bleeding.No other injuries were reported to occur.The reporter was not sure if the bleeding experienced by the patient was related to the reported stapler issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFORM
Common Device Name
STAPLER 60 RELOAD BLACK
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14847888
Report Number2955842-2022-12516
Device Sequence Number20032111
Product Code GDW
UDI-Device Identifier10886874113858
UDI-Public(01)10886874113858(10)N/A
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K173721
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other,Notification
Type of Report Initial
Report Date (Section B) 01/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48360T-08
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2020
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/14/2020
Initial Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
ISIFA2022-02-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient Age66 YR
Patient SexFemale
Patient Weight115 KG
-
-