Model Number 3525-1360S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Implant Pain (4561)
Date of Event 06/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "patient felt crack in left hip while walking and strong pain after that.Event happened after surgery, x ray was made.Revision surgery will be made on (b)(6) 2022.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the nail was completely broken in the thinned webs of the proximal drill hole for the lag screw.The appearance of the breakage surfaces identified the nail had broken in fatigue manner ¿ indicated by lines of rest and missing plastic deformation.The incipient crack had started close to the smallest cross section at lateral in the anterior web and had progressed towards medial.In the lateral area the edge had become damaged by most likely a drill.This was caused intra-operatively by contact with the step drill while preparing the cannulation for the lag screw.The resulting edge had weakened the material significantly.During and after the breakage nail and lag screw had been significantly loaded ¿ indicated by significant material abrasion in the medial, inferior edge of the proximal drill hole.Additional overlying stresses in the area nail -/ lag screw-edge contributed to the origin of the incipient crack.Dimensional inspection indicates that the device is within the specifications.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance with the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
As reported: "patient felt crack in left hip while walking and strong pain after that.Event happened after surgery, x ray was made.Revision surgery will be made on (b)(6) 2022.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, LEFT GAMMA3® Ø11X360MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14847988
Report Number0009610622-2022-00313
Device Sequence Number4832497
Product Code HSB
UDI-Device Identifier07613252274285
UDI-Public07613252274285
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2026
Device Model Number3525-1360S
Device Catalogue Number35251360S
Device Lot NumberK0BE107
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer10/18/2022
Initial Report FDA Received Date06/28/2022
Supplement Report FDA Received Date11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-