Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis deflation.Manufacturer¿s reference number: (b)(4).
|
|
On 19-jan-2023, the mentor failure analysis lab received the device for evaluation.Additionally, an updated implantation date of (b)(6) 2002 was reported.On 24-jan-2023, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced a deflation in the breast implant.The product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and a tear was noted on the posterior view measuring approximately 0.3 cm.A microscopic examination was performed and the cause of the deflation could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.A second product was received.No adverse events were reported for this concomitant (contralateral) device.Therefore, no further analysis is required.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|