• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S FOLYSIL CATHETER; CATHETER, RETENTION TYPE, BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S FOLYSIL CATHETER; CATHETER, RETENTION TYPE, BALLOON Back to Search Results
Model Number AA61061002
Medical Device Problem Codes Break (1069); Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
According to available information, this device required explantation due to failure to deflate.The device was not able to deflate so it was ruptured and removed.No other adverse patient effects were reported.
 
Additional Manufacturer Narrative
A folysil catheter was returned for analysis.The returned device was examined and it was noted the balloon was burst and the stylet was bent.A silicone balloon burst could come from various causes.In this case coloplast can not identify a specific root cause as insufficient information has been provided to do so.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLYSIL CATHETER
Common Device Name
CATHETER, RETENTION TYPE, BALLOON
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
Manufacturer (Section G)
COLOPLAST A/S MANUFACTURING
9 avenue edmond rostand
sarlat-la-caneda 24206
FR   24206
Manufacturer Contact
usaby angela kilian
1601 west river road north
minneapolis, MN 55411
8007880293
MDR Report Key14849400
Report Number9610711-2022-00047
Device Sequence Number4833062
Product Code EZL
UDI-Device Identifier03600040126844
UDI-Public03600040126844
Combination Product (Y/N)N
PMA/510(K) Number
K013174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2026
Device Model NumberAA61061002
Device Catalogue NumberAA6106
Device Lot Number7835654
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/02/2022
Initial Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
Patient SexMale
-
-