NORTHGATE TECHNOLOGIES, INC AUTOLITH TCH 1.9FR PROBE, 375CM DS BSC; INTRA-CORPOREAL ELECTROHYDRAULIC LITHOTRIPSY (IEHL) PROBE Back to Search Results
Model Number 72-00322-0
Medical Device Problem Codes No Apparent Adverse Event (3189); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
Patient undergoing ercp (endoscopic retrograde cholangiopancreatography) cholangioscopy with ehl (electrohydraulic lithotripsy).Four probes were used.Per attending physician: first probe broke outside of patient after use.Second probe used until it no longer worked.Third probe tip broke off.Probes #2 & #3 are from the same lot #.
 
Additional Manufacturer Narrative
On 29jun2022, northgate technologies inc received three separate fda 3500a report filing notifications against 3 probes involved in likely the same event as they were from the same institution, with same event date and same event description.We initiated nc (b)(4) for the investigation and reached out to the distributor for information pertaining to complaints from the end user as northgate technologies, inc was not aware of the malfunctions prior to the medwatch reporting.On 04jul2022, the distributor reported the three complaints to nti, against the fda 3500a reports submitted by the end user.The distributor indicated there was no health consequences or impact to patient outcome.On 01sep2022, the distributor has closed the complaint after three failed attempts requesting for product return with the end user.The distributor informed nti to close the complaint as no product return.Capa22073 was closed with assessment of available information as a physical evaluation of the product was unable to be performed.No patient death or serious injury was reported.The device history record for bsc15867 from february 2022 (mo 15867) was reviewed and the lot passed all testing.The nor-doc-dra-0025 risk analysis was reviewed.Risk id 7.2.12(a) and (b) refers to degradation leading to particle remaining in the body that may cause blockage possibly from the tip detaching, or firing the probe past useful life may cause firing inside the polyimide sheath or probe deterioration causing a procedure delay.Product labeling indicates "single-use only" and a lifetime expiration date.Generator tracks probe life and alerts operator as the probe comes close to and when it reaches end of life.Operator's manual warns to remove all particles.The tips can be removed by minimally invasive means or the tips will pass through the patients system with no harm to the patient.Risk id 7.3.5 (d) refers to accidental mechanical damage that leads to "no spark due to damaged probe - procedure delayed".Per the operator's manual on page 2 under warnings "to avoid procedural delays and possible incomplete stone fragmentation, please have two (2) or more ehl probes available for each lithotripsy procedure." the risk of death or serious injury from the stated issue is remote.A follow up report will be submitted if further information becomes available or product is returned back to nti for evaluation.
 
Event or Problem Description
Event title: confidential.Describe the event or problem: patient undergoing ercp (endoscopic retrograde cholangiopancreatography) cholangioscopy with ehl (electrohydraulic lithotripsy).Four probes were used.Per attending physician: first probe broke outside of patient after use.Second probe used until it no longer worked.Third probe tip broke off.What was the original intended procedure? : ercp.What problem did the user have (check all that apply) :device failed (e.G.Broke, couldn't get it to work or stopped working) ; device malfunction - that is, the device did not do what it was supposed to do.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOLITH TCH 1.9FR PROBE, 375CM DS BSC
Common Device Name
INTRA-CORPOREAL ELECTROHYDRAULIC LITHOTRIPSY (IEHL) PROBE
Manufacturer (Section D)
NORTHGATE TECHNOLOGIES, INC
1591 scottsdale court
elgin IL 60123
MDR Report Key14851262
Report Number14851262
Device Sequence Number14048887
Product Code FFK
UDI-Device Identifier00817183020448
UDI-Public00817183020448
Combination Product (Y/N)N
Initial Reporter StateTX
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 10/07/2022
Report Date (Section F) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/03/2025
Device Model Number72-00322-0
Device Catalogue NumberM00546620
Device Lot NumberBSC15867
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age4 MO
Event Location Hospital
Date Report to Manufacturer06/29/2022
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date10/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
Patient Weight65 KG
Patient EthnicityHispanic
Date Report Sent to FDA06/14/2022
-
-