Corrected information: d9, h3: the complaint device was not returned to cook for investigation.Summary of event: as reported, a flexor raabe guiding sheath "broke" when it was pulled.The provided photo shows the hub separated from the sheath.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures due to the occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation; however, a photo provided by the customer shows that the hub separated from the sheath.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu states ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.¿ the ifu also instructs the user to avoid application of traction to the hub during removal.The information provided upon review of the dmr, dhr, and ifu suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a component failure contributed to this failure mode.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|