(b)(4).Additional information received: was surgery delayed due to the reported event? unknown, was procedure successfully completed? unknown, were fragments generated? unknown, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences? no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study? unknown.Additional information has been requested however not received.Attempts have been made to obtain the device, however not received to date.If further details are received at a later date a supplemental medwatch will be sent.Did the needle fall into the patient? if yes, was the needle piece(s) retrieved during the same procedure? were x-rays taken to locate the needle piece(s)? what measures were taken to retrieve the broken piece? was there any additional tissue damage as a result of searching for the needle piece? does a piece of the needle remain in the patient¿s tissue? if yes, are any plans in place to remove the needle piece in future? what tissue structure the broken needle was located? what is the surgeon's opinion of consequences to the patient? device return status: evaluation: a photo of the product has been received.No product has been received.This is an analysis of a photo submitted for evaluation.During the visual analysis the following was observed: the photograph shows two boxes of closed monocryl product y417h.Image is not clear to determine failure mode or reported condition.Based on the set of photos review, the event describe is not confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.We value the opportunity to fully analyze the instrument upon its return.As part of quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint #(b)(4).Date sent to the fda: 8/3/2022.H3 investigational summary: three suture needles were submitted fractographic evaluation.The components were identified as product code y417h.It was requested that assess the fracture mode of the failures.A fracture was observed at the tip of one needle.One side of the needle was received, the mating fracture surface was not provided for this evaluation.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the tip of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.In order to avoid this kind of damage grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|