| Model Number |
QUANTUM 4FRONT |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Bruise/Contusion (1754)
|
| Date of Event |
06/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Provider alleges the caster stem snapped while the patient was in the chair allegedly causing him to fall out of the chair.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
|
| |
|
Event or Problem Description
|
|
Provider alleges the caster stem snapped while the patient was in the chair allegedly causing him to fall out of the chair.
|
| |
|
Additional Manufacturer Narrative
|
|
(based upon available information regarding the alleged incident at the time of the evaluation): caster fork complete & caster arm only the caster stem was fractured about 3/4" from the nylock nut.The extreme abrading to the caster stem and arm would suggest that the caster pivot was loose/compromised for
some time before the fracturing.This condition would have been noticeable to the user.As the monthly check section of the owner's manual references caster function and
hardware, the likely root cause of the fracturing was improper care & maintenance.
|
| |
|