ASCENSION ORTHOPEDICS, INC. IMPLANT IMPACTOR Back to Search Results
Model Number IMP30000
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a mrh surgery, an mrh head impactor broke into two pieces at the handle during impaction of implant.Both pieces were recovered, it was confirmed that no pieces were left inside of the patient via mini c-arm.Surgery was resumed, without any delay, with the same device.Patient was not harmed as consequence of this problem.
 
Additional Manufacturer Narrative
Section h3, h6: the product has not been received at the aus site for evaluation and photographs were not provided, so the complaint could not be confirmed.The following investigative actions were performed.Refer to the respective tasks for further detail.Complaint history review: the complaint history review identified no similar reported events for this device or for the same product lot.No adverse trend was identified.Future complaints for this failure mode will continue to be monitored and investigated as required.Risk management review (device): identified no issues which could have caused or contributed to the reported event.The failure mode was previously identified by the risk management file and the anticipated risk level is acceptable.Capa/nc/pra/hhe/field action review: identified no previous events or issues which could have caused or contributed to the reported event.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Device labeling/ifu review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met labeling requirements upon release for distribution.Product prints/specifications/procedure review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met product specifications upon release for distribution.As the device has not been received for evaluation, a definitive root cause could not be determined.The complaint alleges no injury to the patient nor a delay during surgery.It was determined that the device likely contributed to the reported event as the most probable cause of the failure was wear and damage due to normal use.The mrh implant impactor is a reusable instrument expected to be used across multiple surgeries.Further, the instrument is used to protect the implant during impaction and is subjected to strong impaction forces during use.Exposure to repeated or excessive force during impaction could cause the instrument to fail with prolonged normal use.Based on this investigation, the need for corrective action is not indicated as no non-conformances or product deficiencies were identified and the risk level is acceptable.If additional information is later received, the complaint may be reopened.No further investigation is warranted for this complaint, though future complaints will be monitored and investigated as required.This investigation can be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT IMPACTOR
Common Device Name
IMPACTOR
Manufacturer (Section D)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer (Section G)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14852595
Report Number3002788818-2022-00043
Device Sequence Number17351779
Product Code HWA
UDI-Device Identifier10381780245728
UDI-Public10381780245728
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIMP30000
Device Catalogue NumberIMP-300-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer08/26/2022
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date08/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-