• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC DC EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC DC EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number DC EXTENSION CABLE
Medical Device Problem Codes Break (1069); Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure, dc extension cable would not transmit power from power supply to hemopro vh-3500.Another cord was used and worked as specified.Upon inspection it was noted that the cord had come apart at the green plug.The case was completed without further incident and there was no patient related event.
 
Additional Manufacturer Narrative
Trackwise id (b)(4).
 
Additional Manufacturer Narrative
Trackwise # (b)(4).Updated section : b4, g4, h2, h3, h6, h10 the reported device is an oem device, therefore, a lot history review was not applicable.A lot number was not provided and the specific product lot number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.H3 other text : device not returned.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DC EXTENSION CABLE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key14852714
Report Number2242352-2022-00577
Device Sequence Number20019533
Product Code GEI
UDI-Device Identifier00607567700888
UDI-Public00607567700888
Combination Product (Y/N)N
Initial Reporter StateAZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDC EXTENSION CABLE
Device Catalogue NumberVH-3030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-