The hakim periton catheter (823045) was not returned for evaluation (as per customer, product not available) and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined; however, the possible root cause for the for the failure issue reported by the customer could be due to the silicone being cut/torn when implanted, or biological debris interfering with the device, as the device was not returned this could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
|
3 of 3 reports.Other mfg report numbers: 3013886523-2022-00295, 3013886523-2022-00296.A physician reported a certas valve (828804) was implanted via v-p shunt on (b)(6) 2017 with unknown setting due to sah (subarachnoid hemorrhage).The valve was implanted with long hakim ventricular catheter (id ns9001) and hakim periton catheter.An unknown failure was suspected; therefore, the valve and catheters were removed and replaced on (b)(6) 2020.It is unknown if the patient experienced any signs and symptoms.
|