Model Number DFT415
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Blurred Vision (2137); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non healthcare professional reported that following implantation of an intraocular lens (iol) the vision has been blurry from first day.The patient can not use out of it.Additional information was requested and received stating as per patient, the lens is not seated right, her eyes are still blurry , her physician tried to correct the iol position , however it didn't help and her eyes are still blurred.Additional information has been received stated that eye still have swelling under right eye as well as right cheek.
 
Event or Problem Description
Additional information has been received stated that the lens explanted from the eye in a secondary procedure and was replaced with a toric lens.No further information available.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14853231
Report Number1119421-2022-01414
Device Sequence Number20060760
Product Code POE
UDI-Device Identifier00380652422062
UDI-Public00380652422062
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/27/2026
Device Model NumberDFT415
Device Lot Number15194034
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/07/2022
Supplement Date Received by Manufacturer07/15/2022
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-