| Model Number |
GIF-XTQ160 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During inspection and testing, service also found the up/down knob could not be locked securely due to wear of the lock engagement lever, the color ring was cracked, the grip was shaved, the forceps channel port was shaved, the switch box was scratched, the right/left knob was scratched, the scope connector cover was scratched, the electrical connector was discolored due to water leakage, the image guide protector was damaged, and the connecting tube was cut.The reported issue (damaged sheath) was not confirmed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
|
| |
|
Event or Problem Description
|
|
The customer reported the sheath of the subject device was damaged.The reported issue was noticed after an egds (endoscopic hemostasis for variceal hemorrhage) procedure.The intended procedure was successfully complete using the subject device.The device had been inspected prior to the intended procedure and no anomalies were observed.There was no patient or user injury reported due to the event.The device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, service found there was foreign material lodged at the distal end area, in the instrument channel and the auxiliary water channel (j-channel).In addition, the same foreign material was found on the plastic distal end cover (c-cover) and the bending section cover (a-rubber).The connecting tube was sticky due to the same foreign material.This report is being submitted for the malfunction found during evaluation of the device (foreign material).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to correct information provided in the initial report.The "date received by manufacturer" should be may 27, 2022.Investigation activities have been opened to manage the actions related to this report and any required mdr reporting.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been over 16 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the foreign material was medical glue which was used by the user and adhered to the device.However, the specific root cause could not be determined at this time.Users may prevent the event by handling the device in accordance the instructions for use (ifu): "do not withdraw the instrument from the endoscope while the needle is extended.This could damage the endoscope or instrument.Avoid aspirating solid matter or thick fluids; instrument channel, suction channel, or suction valve clogging can occur.If the suction valve clogs and suction cannot be stopped, disconnect the suction tube from the suction connector on the endoscope connector.Turn the suction pump off, detach the suction valve, and remove solid matter or thick fluids." olympus will continue to monitor field performance for this device.
|
| |
|