AMO MANUFACTURING USA, LLC WHITESTAR SIGNATURE PRO; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number NGP680301
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Telephone number: (b)(6).Through follow up with customer it was reported that physician was unhappy with the results after surgery and had complained that the equipment was more aggressive to the corneas than normal in comparison to the regular signature vs signature pro.Device evaluation: the engineer went on site to evaluate the system and could not find any issue.The system meet all specifications.After he spoke with the nurse it appears that the wrong program was used during the surgery.Manufacturing records review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a vitrectomy was performed due to a rupture of the capsular bag that motivated the migration of the vitreous to the anterior chamber.The patient is well and stable in terms of vision.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITESTAR SIGNATURE PRO
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14854361
Report Number3012236936-2022-01527
Device Sequence Number4908917
Product Code HQC
UDI-Device Identifier05050474596146
UDI-Public(01)05050474596146
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K151636
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberNGP680301
Device Catalogue NumberNGP680301
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Initial Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;