• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Death
Event or Problem Description
The customer contacted stryker to report that their device was unable to read the patient's ecg through the paddles lead and patient ecg leads.Without this functionality the device would not have the ability to function in aed mode, and the need for defibrillation therapy may not be determined.As a result defibrillation therapy may be delayed or would not be available if needed.A backup device was used and when connected to the defibrillation electrodes already attached to the patient, the ecg was present.The patient did not survive.
 
Additional Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.The customer informed stryker that no further patient information is available.Patient fields in which information is not provided were intentionally left blank.Stryker performed an initial evaluation of the customer¿s device and was unable to duplicate the reported issue.Stryker performed a clinical review of the reported issue and determined there was insufficient data to determined the impact of the device use, therefore it is unknown if the device caused or contributed to the reported outcome.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Additional Manufacturer Narrative
Stryker further examined the customer¿s device and was unable to duplicate the reported issue.Stryker download the electronic records from the device and was able to verify the reported issue.As a pre-caution stryker replaced the system and therapy pcb.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.
 
Event or Problem Description
The customer contacted stryker to report that their device was unable to read the patient's ecg through the paddles lead and patient ecg leads.Without this functionality the device would not have the ability to function in aed mode, and the need for defibrillation therapy may not be determined.As a result defibrillation therapy may be delayed or would not be available if needed.A backup device was used and when connected to the defibrillation electrodes already attached to the patient, the ecg was present.The patient did not survive.
 
Additional Manufacturer Narrative
Section b1 of the supplemental medwatch report (mfr report #0003015876-2022-01341-001).Indicates: adverse event/product problem.Product problem.Outcomes attributed to ae - blank.Section b1 of the supplemental medwatch report (mfr report #0003015876-2022-01341-001) should indicate: adverse event/product problem - adverse event and product problem.Outcomes attributed to ae - death.Section h1 of the supplemental medwatch report (mfr report #0003015876-2022-01341-001) indicates: type of reportable event - malfunction.Section h1 of the supplemental medwatch report (mfr report #0003015876-2022-01341-001) should indicate: type of reportable event - death.
 
Event or Problem Description
The customer contacted stryker to report that their device was unable to read the patient's ecg through the paddles lead and patient ecg leads.Without this functionality the device would not have the ability to function in aed mode, and the need for defibrillation therapy may not be determined.As a result defibrillation therapy may be delayed or would not be available if needed.A backup device was used and when connected to the defibrillation electrodes already attached to the patient, the ecg was present.The patient did not survive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key14854521
Report Number0003015876-2022-01341
Device Sequence Number5562814
Product Code MKJ
UDI-Device Identifier00883873871669
UDI-Public00883873871669
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001960
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2022
Supplement Date Received by Manufacturer12/27/2022
01/17/2023
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date12/27/2022
01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-