Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The procedure has not been rescheduled to date.Multiple, unsuccessful attempts were made to obtain more information from the customer.The device was not returned for evaluation.There was no allegation of device failure.Risk of pneumothorax is documented in (b)(4), rev f, monarch bronch risk management report and is deemed to be acceptable.Based on the information available for this complaint, the root cause of the reported event is related to a known inherent risk of the device and procedure as documented in the labeling.
 
Event or Problem Description
It was reported that during a monarch-assisted bronchoscopy procedure, the patient experienced a pneumothorax.The target location was next to a fissure.A chest tube was placed.The procedure has not been rescheduled to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH PLATFORM
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES
Manufacturer (Section D)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA 94065
Manufacturer (Section G)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA 94065
Manufacturer Contact
harminder kaur
150 shoreline drive
redwood city, CA 94065
MDR Report Key14856504
Report Number3014447948-2022-00010
Device Sequence Number17533327
Product Code EOQ
UDI-Device IdentifierB634MON00000501R0
UDI-Public+B634MON00000501R0/16D20210917
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K193534
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Initial Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-