• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P TAPERFILL HIP; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 13

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P TAPERFILL HIP; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 13 Back to Search Results
Model Number 425-97-013
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870)
Date of Event 06/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as fall.The previous surgery and the surgery detailed in this event occurred 2 months and 1 week apart.The healthcare professional indicated there was a serious risk to the patient.There no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was a nonconformance associated with the main part #425-97-013, taperfill hip stem, lateral offset, size 13 which documents that out of 12 parts lot, 1 part was rejected and scrapped on router.All other items in the lot were met with the design, fit and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to a fall.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Agent has clearly mentioned that the patient fell.It seems that the event might have possibly occurred due to, improper surgical technique and incorrect implant selection, patient noncompliance with medical instructions, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - due to fall and broke femur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERFILL HIP
Common Device Name
TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 13
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14856590
Report Number1644408-2022-00858
Device Sequence Number14751340
Product Code LPH
UDI-Device Identifier00888912078719
UDI-Public(01)00888912078719
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number425-97-013
Device Catalogue Number425-97-013
Device Lot Number178R2106
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2022
Initial Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-402 LOT 871B1357
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age72 YR
Patient SexMale
-
-