| Model Number |
309-10-50 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was damaged, fractured locking clip was damaged during trailing of device.No parts or pieces fell into the patient wound site.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.No patient information/medical history available.Device is returning.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the broken instrument reported was likely the result of not considering specific performance requirements to capture user needs and design inputs (material durability during reprocessing/sterilization as well as force applied over lifetime of device), which led to insufficient testing to demonstrate continued performance over time and resulted in fracture during surgical use.
|
| |
|