Model Number 309-10-50
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was damaged, fractured locking clip was damaged during trailing of device.No parts or pieces fell into the patient wound site.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.No patient information/medical history available.Device is returning.
 
Additional Manufacturer Narrative
Pending evaluation.
 
Additional Manufacturer Narrative
Section h10: (h3) the broken instrument reported was likely the result of not considering specific performance requirements to capture user needs and design inputs (material durability during reprocessing/sterilization as well as force applied over lifetime of device), which led to insufficient testing to demonstrate continued performance over time and resulted in fracture during surgical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key14859523
Report Number1038671-2022-00745
Device Sequence Number5074124
Product Code LXH
UDI-Device Identifier10885862267726
UDI-Public10885862267726
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number309-10-50
Device Catalogue Number309-10-50
Device Lot Number84967003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-