| Model Number |
15 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.Stryker evaluated the customer's device and was unable to duplicate the reported issue but was able to verify the reported issue was logged in the device¿s memory.After completing other unrelated repairs, proper device operation was observed through functional and performance testing and the device returned to the customer for use.The cause of the reported issue could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly turned off.In this state the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however there was no adverse event reported.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly turned off.In this state the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker received additional patient information from the customer.The customer provided stryker with the available patient information.Patient fields in which information is not provided were intentionally left blank.
|
| |
|