Model Number 013554
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during an acl reconstruction, the tip of the tendon stripper broke off inside the patient's thigh when the surgeon was trying to harvest the graft.The broken pieces were retrieved from the patient using tweezers.The procedure was completed with a non-significant delay using a smith and nephew back-up device.No further complications were reported.Patient status is fine.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
B2 was updated.H3, h6: the reported device was received for evaluation.A visual inspection revealed the device was returned outside of original packaging.The device is coated in debris with signs of wear.The batch number and product number are visible.The distal tip shows a fracture and the detached piece is returned in a cup of solution.A review of the customer provided image revealed the distal tip of the device has become fractured.Undated photos of the device were provided for review and confirms the reported breakage.There was a relationship found between the device and the reported event.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors, which could have contributed to the complaint event, include an application of unintended inappropriate or excessive force to the device.Per case details, the broken pieces were retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDON STRIPPER SLOTTED
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14860129
Report Number1219602-2022-00927
Device Sequence Number14763893
Product Code LXH
UDI-Device Identifier03596010091093
UDI-Public03596010091093
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number013554
Device Catalogue Number013554
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date09/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-