• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL, SPIDER 2, TENET 7615; TABLE, OPERATING-ROOM, ELECTRICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL, SPIDER 2, TENET 7615; TABLE, OPERATING-ROOM, ELECTRICAL Back to Search Results
Catalog Number 72203299S
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during a shoulder scope, the tenet spider's arm was not holding.Upon connecting the patients arm to the device, the hydraulic failed.The patient's arm was not injured.The device was re-draped in a sterile fashion, but the arm failed again in the same manner.The procedure was completed without a delay, by wrapping sterile coban around the locking mechanism, and sterile personnel supporting the spider arm through the duration of the procedure.No further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.There was a relationship found between the returned device and the reported event.A visual inspection was performed on the product.The device was received pressurized.No accessories were included.There was pooling oil noted at the distal ball.A functional evaluation was performed.The device failed the weight test.The piston migration was at 1.8 inches.The distal seals appeared to be slightly compressed.The reported failure was confirmed and the root cause has been associated with a seal failure.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately.There were no indications to suggest the anticipated risk is not adequate.It is recommended that the spider 2 ifu be reviewed regarding depressurizing the spider 2 for storage to prevent the premature deterioration of the seals, pressure cups and pressure rings.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL, SPIDER 2, TENET 7615
Common Device Name
TABLE, OPERATING-ROOM, ELECTRICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14860614
Report Number1643264-2022-00228
Device Sequence Number5195523
Product Code GDC
UDI-Device Identifier00885554029673
UDI-Public00885554029673
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72203299S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date06/29/2022
Supplement Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-