U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number BMK6F105BER130
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure in the left internal carotid artery (ica) using a benchmark 6f 071 delivery catheter (benchmark), a non-penumbra microcatheter and a non-penumbra sheath.After completion of the procedure, while retracting the benchmark, the benchmark started to unravel.Subsequently, the physician removed the benchmark by pulling it out.The procedure was completed at this point.It was reported that upon inspection of the benchmark next day, the proximal end of the benchmark was noticed to be separated with the coil wind attached to the rest of the catheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned benchmark confirmed that the catheter was fractured.If the benchmark is manipulated at an extreme angle during use, damage such as a kink may occur.Further manipulation of a kinked device may worsen to a fractured.Upon retraction of a fractured device, unraveling may occur.Further evaluation revealed bends throughout the length of the catheter shaft.This damage was incidental to the reported complaint and may have occurred during packaging for return.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK 6F 071 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key14860816
Report Number3005168196-2022-00308
Device Sequence Number5195563
Product Code DQY
UDI-Device Identifier00814548014029
UDI-Public814548014029
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142321
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/29/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/29/2024
Device Catalogue NumberBMK6F105BER130
Device Lot NumberF108073
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2022
Initial Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
Date Report Sent to FDA01/01/2005
-
-