| Model Number |
SSLP-190 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture (baker grade iii)" against left device.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified, weight to the spec, white particles on the shell.After autoclave cycle was observed: cloudy color in the gel.The device is considered intact and functional.Based on the device analysis the final assessment is: intact and functional.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture (baker grade iii)" against left device.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture (baker grade iii)" against left device.Device was explanted.
|
| |
|
Search Alerts/Recalls
|