• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH ANN, PROXIMAL HUMERUS, RIGHT, 9X160MM; AFFIXUS NATURAL NAIL SYSTEM HUMERAL NAIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH ANN, PROXIMAL HUMERUS, RIGHT, 9X160MM; AFFIXUS NATURAL NAIL SYSTEM HUMERAL NAIL Back to Search Results
Model Number N/A
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: blunt tip screw, a 4x44mm item# 47-2486-044-40 lot# 3024704, blunt tip screw, a 4x46mm item# 47-2486-046-40 lot# 3062600, blunt tip screw, a 4x48mm item# 47-2486-048-40 lot# 3006014, torque limiting handle item# 27923 lot#unknown.Report source: (b)(6).Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2022-00349, 0009613350-2022-00350, 0009613350-2022-00351.
 
Event or Problem Description
It was reported that the patient underwent a revision procedure of the right hip approximately six (6) months post implantation due to migration.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.As no product was returned, visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Devices are used for treatment.No surgical report or x-rays are available for investigation.Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition, behavior or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.The investigation also considered whether the design limitation of the corelock mechanism could be a potential contributing factor.However, further biomechanical testing identified that the product performance is in an acceptable range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANN, PROXIMAL HUMERUS, RIGHT, 9X160MM
Common Device Name
AFFIXUS NATURAL NAIL SYSTEM HUMERAL NAIL
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14869515
Report Number0009613350-2022-00352
Device Sequence Number5139127
Product Code HSB
UDI-Device Identifier00889024505780
UDI-Public(01)00889024505780(17)301130(10)3050610
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181827
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47-2496-160-09
Device Lot Number3050610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date09/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-