Model Number IEHA466
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie dental complaint (b)(4).
 
Event or Problem Description
It was reported that the encode healing abutment at end of shaft the tip is flared out.This defect would not allow it to seat.Lab had to had grind the encode healing abutment to make it the correct shape to fit.Tooth site # 13.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.The following sections are being reported: d4: expiration date and unique identifier (udi) number.H2: if follow-up, what type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: type of investigation.H6: investigation findings.H6: investigation conclusions.H10: additional narratives/data.The product was not returned.The reported event was confirmed via customer's.Picture.The coob could not be confirmed.Based on the evaluation, device malfunction has occurred.However, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported device/packaging that did or could cause or contribute to the reported event.Monthly post market trending review identified no actionable trends or corrective actions for the reported event or device.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the product/packaging was within specifications and conforming when it left zimmer biomet.Dhr review for the lot had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.All products were conforming at the time they left zimmer biomet.Lot was inspected and passed all acceptance criteria by qa.The dhr was further reviewed to verify the product release information.No malfunction or nonconformance was identified.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.A definitive root cause could not be determined.No further investigation or immediate capa/hhe/d escalation is required, as the complaint investigation did not confirm the product/packaging was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
 
Event or Problem Description
No additional or corrected information to report.
 
Event or Problem Description
No additional or corrected information to report.
 
Additional Manufacturer Narrative
This report is being submitted to report corrected information.Further review of the investigation confirmed that the alleged seating failure could not be verified, and malfunction could not be verified.The following sections are being reported: h2: if follow-up, what type, h6: investigation findings, h6: investigation conclusions, h10: additional narratives/data.The product was not returned.The reported event was non-verifiable.Based on the evaluation, device malfunction could not be verified.Additionally, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported device/packaging that did or could cause or contribute to the reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN® BELLATEK® ENCODE® HEALING ABUTMENT 4.1MM(D) X 6MM(P) X 6MM(H)
Common Device Name
HEALING ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14873789
Report Number0001038806-2022-01014
Device Sequence Number20568840
Product Code NHA
UDI-Device Identifier00844868004691
UDI-Public(01)00844868004691(17)270503(10)1254351
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K072642
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIEHA466
Device Catalogue NumberIEHA466
Device Lot Number1254351
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer12/15/2022
01/03/2023
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date12/15/2022
01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-