• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOFOCUS HEARTLIGHT; HEARTLIGHT X3 CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOFOCUS HEARTLIGHT; HEARTLIGHT X3 CATHETER Back to Search Results
Model Number 18-5000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nerve Damage (1979)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No device deficiency reported.Phrenic nerve injury leading to diaphragmatic paralysis is a known potential adverse event for catheter ablation procedures disclosed in product labeling.
 
Event or Problem Description
During a pulmonary vein isolation (pvi) procedure to treat atrial fibrillation, phrenic capture was lost while ablating the right superior pulmonary vein (rspv).At the conclusion of the case the phrenic nerve could not be paced.Decreased excursion of the right hemidiaphragm was observed under fluoroscopy.An update three weeks post-procedure found the patient asymptomatic and the physician believed the phrenic nerve injury had resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTLIGHT
Common Device Name
HEARTLIGHT X3 CATHETER
Manufacturer (Section D)
CARDIOFOCUS
500 nickerson rd.
marlboro MA 01752
Manufacturer Contact
ian christianson
500 nickerson rd.
marlboro, MA 01752
MDR Report Key14874169
Report Number1225698-2022-00015
Device Sequence Number20145966
Product Code OAE
UDI-Device Identifier00859373007433
UDI-Public00859373007433
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P150026
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number18-5000
Device Catalogue Number18-5000
Device Lot Number16466-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Initial Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age54 YR
Patient SexFemale
Patient RaceWhite
-
-