Model Number SA60WF
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported after the iol was delivered in the bag, a very clearly marking / a scratch was observed in the middle of the lens.The lens has been cut for removal from the patient¿s eye.The replacement lens was used.No patient harm reported, symptoms resolved.Additional information has been requested.
 
Additional Manufacturer Narrative
The product was returned for analysis and damage was observed to the iol.Iol returned in three pieces in the lens case.Solution is dried on both surfaces of the optic and haptics.One haptic is broken/torn and is returned, the other is intact.The optic is cut in half and scratched/marked-rejectable.The product investigation could not identify a root cause for the reported complaint ¿a very clearly marking / a scratch was in the middle of the lens¿.The product was subject to handling and the reported damage was highlighted post implantation.The returned iol shows evidence of handling by the customer due to the presence of solution dried both surfaces of the optic and haptics and how it was returned in three pieces.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol contributed to the event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14874862
Report Number9612169-2022-00304
Device Sequence Number5727980
Product Code HQL
UDI-Device Identifier00380652332736
UDI-Public00380652332736
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/02/2026
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.235
Device Lot Number25202592
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/03/2022
Supplement Date Received by Manufacturer08/25/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; UNKNOWN OVD