• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number RONYX30015UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
An attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a moderately tortuous, moderately calcified lesion in the ostium of the left main (lm) coronary artery.Excessive force was not used during delivery.It was reported that stent dislodgement occurred during delivery to the lesion.The stent was not removed and was deployed.No patient injury was reported.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: patient history provided.The device was not inspected before use.The device was prepped per ifu.The lesion was pre-dilated.The device did not pass through a previously deployed stent or cell of a stent.The stent was not removed and was deployed in the left main by ballooning with the stent balloon.The patient is reported to be stable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14874919
Report Number9612164-2022-02473
Device Sequence Number20120144
Product Code NIQ
UDI-Device Identifier00643169556980
UDI-Public00643169556980
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2025
Device Model NumberRONYX30015UX
Device Catalogue NumberRONYX30015UX
Device Lot Number0011072289
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2022
Supplement Date Received by Manufacturer06/30/2022
09/07/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date07/18/2022
09/09/2022
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-