| Model Number |
PINNACLE3 SMART ENTERPRISE SERVER |
| Medical Device Problem Code |
Program or Algorithm Execution Failure (4036)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing investigation currently underway.Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: (b)(4).Date of report 20220629.
|
| |
|
Event or Problem Description
|
|
Exported dose from a plan is wrong when imported back into pinnacle and when exporting to 3rd party systems.This behavior was identified by the customer while doing dose comparison testing between software version 18.0.1 and software version 9.8.After exporting plans to 3rd party system mim it was discovered that the dose was different between the two plans.This issue is currently under investigation by the manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s investigation has identified a specific coding issue that was introduced in the development phase of pinnacle software version 16.4.2 as the root cause of this issue.This coding issue has been corrected and this fix is being distributed to the customer via installation of a service pack, specifically in this case, 18.0.1 service pack 1.Internal cross reference: complaint pr# (b)(4).Date of report 20220916.
|
| |
|
Event or Problem Description
|
|
Exported dose from a plan is wrong when imported back into pinnacle and when exporting to 3rd party systems.This behavior was identified by the customer while doing dose comparison testing between software version 18.0.1 and software version 9.8.After exporting plans to 3rd party system mim it was discovered that the dose was different between the two plans.
|
| |
|