HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Pumping Problem (3016); Noise, Audible (3273)
|
Patient Problem
Thrombosis/Thrombus (4440)
|
Event Date 07/28/2021 |
Event Type
Injury
|
Event Description
|
A journal article was reviewed that contained information regarding detection of thrombi.The article reports a patient who was admitted emergently with recurrent left ventricular assist device (vad) low flow alarms.Vad thrombosis was suspected through logfile readouts which displayed low flows, power spikes, and ¿third-harmony¿ sound indicative of impeller imbalance.A computerized tomography (ct) visualized activity accumulation near and around the inflow cannula and in the area of the outflow graft.Systemic lysis therapy with t-pa was successfully applied under close vital sign monitoring in the intensive care unit (icu).Eventually, the patient was discharged with normalized vad function parameters.The status/disposition of the vad and outflow graft appears to be still in use.No further patient complications have been reported as a result of this event.Further follow up did not yet yield any additional information.
|
|
Manufacturer Narrative
|
Additional products: heartware ventricular assist system ¿ outflow graft, model #: unk/ catalog #: unk/ expiration date: unk / serial or lot#: unknown, udi #: asku, device available for eval: no, mfg date: unk, labeled for single use: no, (b)(4).This information is based entirely on journal literature.Medtronic was made aware of this event through a search of literature publications.This event occurred outside the us.Patient information is limited due to confidentiality concerns.The event only occurred with one patient but specific details on the patient were not provided.The model listed in the report is a representative of the model family, as there is no specific model listed.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.A request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: gmp-compliant radiosynthesis of [18f]gp1, a novel pet tracer for the detection of thrombi.Pharmaceuticals.2021, 14, 739.Doi: 10.3390/ph14080739 investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for investigation and anlysis completion.Product event summary: one (1) pump with unknown serial number was not returned for evaluation.Log file analysis could not be performed since log files covering the reported event date were not available for analysis.The reported high power, low flow, and pump sound events could not be confirmed due to insufficient evidence.The reported thrombus event was confirmed via visual evidence, which revealed evidence of thrombus within the device.Based on the available information, the device may have caused or contributed to the reported event.Based on the risk documentation, possible causes of the reported low flow event may be attributed to multiple factors including but not limited to thrombus at the inflow cannula/outflow graft, constriction at the outflow graft, poor vad filling, and/or inappropriate pump rotational speed.Based on risk documentation, multiple factors may have contributed to the high power event including but not limited to thrombus formation/ingestion, high flows, or incorrect setting of alarm threshold.Based on the risk documentation, the reported pump sound event may be attributed to multiple factors including but not limited to thrombus within the device leading to impeller imbalance or the placement of the pump which allows contact with fixed or rigid anatomical structure.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.One (1) outflow graft with unknown lot number was not returned for evaluation.Review of the controller log files could not be conducted since log files were not available for analysis.Visual evidence provided revealed evidence of thrombus within the device.As a result, the reported thrombus event was confirmed.Based on the available information, the device may have caused or contributed to the reported event.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Additional products: brand name: heartware ventricular assist system ¿ outflow graft h6: fda method code(s): b15, b17 h6: fda results code(s): c19 h6: fda conclusion code(s): d12 investigation of this event is completed, and the file will be closed.If new information is received, the file will be re-opened, and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|