| Catalog Number |
711.88 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample has been received by manufacturing that has not yet been evaluated.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during a vitrectomy procedure three ophthalmic forceps had meshing failure and made a squeaking noise when opening and closing during insertion.The surgery was performed and completed after replacing the product with another one.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is completed sample-1.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample is received with inner blister, outer blister and without cover foil.The sample shows macroscopic signs of damage.The shaft is bent.The sample shows signs of surgery residues and is returned in a opened status.The sample was visually inspected with the aid of a photomicroscope with various magnifications.The sample was visually inspected and functionally tested.It was noticed that the shaft of the device is bent, there was a squeaking noise to hear, but the sample was able to open and close normally although the shaft was bent.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.The customers complaint is confirmed it cannot be determined how or where the damage to the sample occurred; therefore a root cause for the customers reported event could not be determined.However, the damage was consistent with damage that can occur due to improper handling.The exact root cause for the customers reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|