BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO XL; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00547110
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro xl retrieval balloon was intended to be used in the common bile duct (cbd) for an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During preparation, it was noted that the shape of the balloon was irregular, which was confirmed via a photo submitted by the customer.The procedure was completed with another extractor pro xl retrieval balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Additional Manufacturer Narrative
Block h6: device code a0406 captures the reportable event of balloon irregular shape.Block h10: investigation results a visual examination of the returned complaint device found that the balloon and the catheter of device had no damages.Functional analysis was performed, and the balloon was inflated without a problem and held pressure.Dimensional examination of the balloon concentricity was performed and was within specification.No other problems with the device were noted.The reported event of balloon's shape crumple was not confirmed.The complaint event description, device analysis and including the product record review results, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.Therefore, the most probable root cause cannot be determined as no problem was detected.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro xl retrieval balloon was intended to be used in the common bile duct (cbd) for an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During preparation, it was noted that the shape of the balloon was irregular, which was confirmed via a photo submitted by the customer.The procedure was completed with another extractor pro xl retrieval balloon.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR PRO XL
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14876756
Report Number3005099803-2022-03551
Device Sequence Number4996569
Product Code FGE
UDI-Device Identifier08714729790402
UDI-Public08714729790402
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K102082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2024
Device Model NumberM00547110
Device Catalogue Number4711
Device Lot Number0028862813
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer09/30/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-