SMITH & NEPHEW, INC. PROBE,SAPHYRE II, 90DEG,3MM,SUCTION; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7210111
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, the insulation rubber of the saphyre ii probe fell off.The fallen off rubber was removed by suction.The procedure was completed using a back-up device.It is unknown if there was a delay.No further complications were reported.
 
Event or Problem Description
It was reported that during an arthroscopy, the insulation rubber of the saphyre ii probe fell off.The fallen off rubber was removed by suction.The procedure was completed with non-significant surgical delay and was finished with a smith and nephew back up device.No further complications were reported.
 
Additional Manufacturer Narrative
H10, h6: a device deficiency was identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that two undated photos where provided for review and confirm the reported insulation rubber removal.Per case details, the rubber from the wand was retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROBE,SAPHYRE II, 90DEG,3MM,SUCTION
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14876856
Report Number3003604053-2022-00022
Device Sequence Number12220410
Product Code GEI
UDI-Device Identifier03596010481498
UDI-Public03596010481498
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2024
Device Model Number7210111
Device Catalogue Number7210111
Device Lot Number2070677
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2022
Supplement Date Received by Manufacturer07/07/2022
06/01/2023
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date07/11/2022
06/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age53 YR
Patient SexFemale
Patient Weight68 KG
-
-