ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0¿, 63 CM LENGTH, 8.5 F; INTRODUCER, CATHETER
|
Back to Search Results |
|
Model Number 407451 |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/09/2022 |
Event Type
malfunction
|
Event Description
|
During the procedure, an air leak was noted following insertion into the patient.When reflux was performed, a bubble appeared.The sheath was exchanged and the procedure was successfully completed with no adverse patient consequences.
|
|
Manufacturer Narrative
|
One 8.5f swartz braided introducer sheath and dilator were received for evaluation.Functional testing confirmed an air leak in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak remains unknown.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
|
|