| Model Number |
SN6CWS |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
05/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.A photo was provided which shows a yellow lens placed inside a the barrel of an ovd syringe.A note is visible in the photo with the patient information hand written on it.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported following an intraocular lens (iol) implant procedure, the patient complained several times postoperatively about poor visual acuity.As a result, surgeon decided to exchange the iol for the same type of lens and same diopter.The surgeon also suspects an irregularity on the anterior surface of the lens.Postoperative result was satisfactory.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned in a surgical syringe filled with a clear liquid.The lens was removed and cleaned with klrp for further evaluation.One haptic was broken in the gusset area (returned).The optic was cut into two portions, typical of insertion and removal.A power and resolution verification was conducted with the larger portion.The optical resolution is acceptable; lens meets specification for focal length equaling a 18.5 diopter lens.No lens issue observed.The product investigation could not identify a root cause for the reported complaint.The lens was returned.A power and resolution verification was conducted.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|