U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941828500
Medical Device Problem Codes Break (1069); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 90% stenosed target lesion was located in the moderately-severely tortuous and mildly calcified right coronary artery.A 5.00 x 28mm synergy xd drug-eluting stent was advanced for treatment.However, as the physician was driving the stent catheter up the guide, resistance was encountered and the proximal portion of the shaft snapped when reaching a tortuous part of anatomy.Furthermore, multiple attempts were made to cross the lesion.The physician was able to pull and remove it out from the guide.A new synergy xd stent was used to complete the procedure.No patient complications were reported.
 
Additional Manufacturer Narrative
Synergy xd mr us 5.00 x 28mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The stent outer diameter was measured using snap gauge and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along the length of the hypotube shaft, as well as a hypotube break located approximately 41cm distal to the distal strain relief.A visual and tactile examination of the outer and mid-shaft section found a shaft polymer extrusion kink located approximately 21cm proximal to the distal tip.No other issues noted with device.
 
Event or Problem Description
It was reported that shaft break occurred.The 90% stenosed target lesion was located in the moderately-severely tortuous and mildly calcified right coronary artery.A 5.00 x 28mm synergy xd drug-eluting stent was advanced for treatment.However, as the physician was driving the stent catheter up the guide, resistance was encountered and the proximal portion of the shaft snapped when reaching a tortuous part of anatomy.Furthermore, multiple attempts were made to cross the lesion.The physician was able to pull and remove it out from the guide.A new synergy xd stent was used to complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14877269
Report Number2134265-2022-06976
Device Sequence Number14707984
Product Code NIQ
UDI-Device Identifier08714729981282
UDI-Public08714729981282
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2023
Device Model NumberH7493941828500
Device Lot Number0027745999
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer07/25/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER - SYNERGY XD 5.0 X 28MM; GUIDE CATHETER - SYNERGY XD 5.0 X 28MM
Patient SexUnknown
-
-