• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DIREXION FATHOM-16 SYSTEM; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DIREXION FATHOM-16 SYSTEM; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 84592
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a fracture occurred.A direxion microcatheter was selected for use.The lesion was accessed through the femoral artery in the right groin and the device was advanced to the target lesion.During removal of the microcatheter, an abnormality was observed near the tip of the device.There was a split at the tip and it looked damaged and worn out.The procedure was completed with another of the same device.There were no patient complications reported.
 
Event or Problem Description
It was reported that a fracture occurred.A direxion microcatheter was selected for use.The lesion was accessed through the femoral artery in the right groin and the device was advanced to the target lesion.During removal of the microcatheter, an abnormality was observed near the tip of the device.There was a split at the tip and it looked damaged and worn out.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the direxion microcatheter was returned for analysis.The device shaft was microscopically analyzed for any damage.The device showed a fracture at the hub.There were also 2 other fractures located 70.2cm and 70.5cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION FATHOM-16 SYSTEM
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14877579
Report Number2134265-2022-07510
Device Sequence Number17398965
Product Code KRA
UDI-Device Identifier08714729839637
UDI-Public08714729839637
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2023
Device Model Number84592
Device Catalogue Number84592
Device Lot Number0027955416
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-