DEPUY MITEK LLC US 5.5 HEALIX ADVANCE BR 3SUTURE ANCHOR W/ PERMACORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE
|
Back to Search Results |
|
| Model Number |
223131 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Occupation: reporter is a j&j employee.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the affiliate in (b)(6) that during an arthroscopic rotator cuff repair procedure on (b)(6) 2022, it was observed that the tip on the 5.5 healix advance br 3suture anchor w/ permacord device broke off when used after making a pilot hole with an awl.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as approriate.Investigation summary the complaint device was received and evaluated.Upon visual inspection, it could be observed the white paper card with the suture is in place as it should.Finally, the tip of the anchor threads was found damaged with deformed condition.The shaft was found in good condition as well as the inserter tip.A manufacturing record evaluation was performed for the finished device lot number: 7l89273, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.A possible root cause can be attributed to procedural variables, such handling of the device or product interaction during procedure.When trying to insert the anchor, the anchor can lead to deformed and could be get stuck in the inserter if a pilot hole is not created adequate to allow proper anchor placement.However, this cannot be conclusively affirmed.As per ifu, inserting the awl or drill to less than the specified depth, axial misalignment or levering with the anchor upon insertion may result in anchor fracture.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
| |
|