The customer reported to olympus, the uretero-reno fiberscope had black spots and tears near the tip.Upon inspection of the customer¿s returned device it was observed, metal part of the curved tube stuck out of the tube.This report is being submitted for the malfunction found during evaluation.There was no harm or user injury reported due to the event.
|
The subject device has been returned to olympus for evaluation.During evaluation, it was observed, curved rubber was broken, the bending angle was insufficient, poor insertion of the treatment tool was noted in the forceps channel, the curved pipe was damaged, both the curved rubber and the curved tube were damaged, and the watertightness was not maintained due to the breakage of curved rubber.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|