Model Number URF-P6
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the uretero-reno fiberscope had black spots and tears near the tip.Upon inspection of the customer¿s returned device it was observed, metal part of the curved tube stuck out of the tube.This report is being submitted for the malfunction found during evaluation.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
The subject device has been returned to olympus for evaluation.During evaluation, it was observed, curved rubber was broken, the bending angle was insufficient, poor insertion of the treatment tool was noted in the forceps channel, the curved pipe was damaged, both the curved rubber and the curved tube were damaged, and the watertightness was not maintained due to the breakage of curved rubber.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to damage that resulted from forcible twisting operation to the right and left while bending of the bending section was fixed or irregular load applied by abrupt bending operation.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14879814
MDR Text Key302785082
Report Number8010047-2022-11128
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2022
Was the Report Sent to FDA? No
Date Manufacturer Received08/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1263-2022
Patient Sequence Number1
-
-