Model Number 222207
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer in china that during an arthroscopic shoulder stabilization procedure on (b)(6) 2022, it was observed that the anchor and shaft on the 5.5mm healix peek 3-suture anchor w/orthocord device could not be detached from one another.During in-house engineering evaluation, it was determined that the cover handle and the shaft were detached from the handle.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation.However, a photo was provided.Upon visual inspection of the photo, the anchor was still in the shaft and held by the suture.The cover handle and the shaft were detached from the handle; however, it is uncertain that the anchor is stuck due to the device was not returned.The photo do not provide enough evidence to confirm this complaint.Hands on analysis should provide the evidence necessary to determine the root cause.The photo provided does not contain enough evidence to determine why the customer experienced the failure, hands on analysis should provide the required evidence to provide a root cause.A manufacturing record evaluation was performed for the finished device lot number: and no nonconformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5MM HEALIX PEEK 3-SUTURE ANCHOR W/ORTHOCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14880339
Report Number1221934-2022-01994
Device Sequence Number5089384
Product Code MAI
UDI-Device Identifier10886705007554
UDI-Public10886705007554
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number222207
Device Catalogue Number222207
Device Lot Number8L67265
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2022
Initial Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-