| Model Number |
N/A |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On june 17,2022 fujifilm healthcare corporation received a complaint regarding oasis xp 1.2t open mri system.The customer complained about the quality of breast imaging.The patient is a known breast cancer patient and is actively undergoing treatment.The site reported the discrepancy in the size of a lesion that was 75 mm on a non-fujifilm system (scan was done on (b)(6) 2022) vs only 5 mm on the oasis xp 1.2t open mri system (scan was done on (b)(6) 2022).There is no death or serious injury associated with event.
|
| |
|
Additional Manufacturer Narrative
|
|
Fujifilm is investigating the root cause and a supplemental report will be submitted pending investigation results.
|
| |
|
Additional Manufacturer Narrative
|
|
On 07/22/2022, fujifilm healthcare corporation concluded its root cause investigation.There were no device-related issues identified as a result of the investigation.It was concluded that it is likely that the issue was due to user error.The training records were reviewed and determined to be adequate, and retraining was determined to be unnecessary.
|
| |
|