• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SET HF20; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SET HF20; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 109841
Device Problem Inaccurate Information (4051)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2022
Event Type  malfunction  
Event Description
It was reported that an error in translation regarding pediatric patients weight occurred when translating instructions for use from the english language to estonian or baltic counties on a prismaflex set hf20 occurred regarding the pediatric patients weight.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Prismaflex hf20 set has been temporarily approved for use in the us under emergency use authorization eua201769 to provide continuous renal replacement therapy (crrt) to treat low weight (8 kg to 20 kg) and low blood volume patients or patients who have acute renal failure, fluid overload, or both, and who cannot tolerate a larger extracorporeal circuit volume in an acute care environment during the coronavirus disease 2019 (covid-19) pandemic.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual device was not available for investigation; however, a retained sample of instruction for use (ifu) was evaluated.The estonian ifu translated into english states: "prismaflex m60 should not be used in children weighing more than 11 kg.Prismaflex m100 and m150 kits should not be used in patients with body weight exceeding 30 kg.¿ the correct sentence in the english ifu states: "prismaflex m60 should be restricted to children with a body weight greater than 11 kg, and m100/150 sets should be restricted to patients with a body weight greater than 30kg.¿ the reported condition was verified.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SET HF20
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14880504
MDR Text Key299887265
Report Number8010182-2022-00179
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeEN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number109841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/27/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-