| Model Number |
83787 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the coil has come out of the catheter.The target lesion was located in the moderately tortuous and non-calcified internal iliac artery.A 8mm x 20cm.035 interlock cube embolics was selected for use.During delivery of the product, a gutter was created by breaking through the wall of the catheter instead of the tip and the coil has come out of the catheter.The entire catheter was removed, however, two-thirds of the coil protruded from the tip.The procedure was completed with a different device.There were no complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: only the main coil was returned analysis.The main coil was kinked and stretched.No more damages were found.
|
| |
|
Event or Problem Description
|
|
It was reported that the coil has come out of the catheter.The target lesion was located in the moderately tortuous and non-calcified internal iliac artery.A 8mm x 20cm.035 interlock cube embolics was selected for use.During delivery of the product, a gutter was created by breaking through the wall of the catheter instead of the tip and the coil has come out of the catheter.The entire catheter was removed, however, two-thirds of the coil protruded from the tip.The procedure was completed with a different device.There were no complications reported.
|
| |
|